Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Fluke 83,85,87 (1 year) Capacitance CAL VER DATE: 20-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 74 STANDARD: Capacitance Decade ============================================================================= # # Source: # Fluke 80 Series Service Manual (PN 834168 April 1989). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification given in the ACC statements. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P J F W 1.002 ASK+ K 1.003 HEAD {CAPACITANCE TESTS} 1.004 STD Capacitance Decade 1.005 DISP Connect the Capacitance Decade and UUT as follows: 1.005 DISP [32] DECADE HI to UUT V Ohm Diode 1.005 DISP [32] DECADE LO to UUT COM 1.006 DISP Rotate the UUT function switch to Ohms and press the 1.006 DISP BLUE button to select Capacitance. 1.007 HEAD CAPACITANCE TESTS: {5.00uF Range} 1.008 DISP Set the Capacitance Decade to 1.00uF. # Note, You must substitute your actual capacitance decade tolerance below. 1.009 ACC 1.00uF 1U 1.010 MEMI Enter UUT reading in microfarads: 1.011 MEME 1.012 MEMC 5 uF 1% 0.02U 2.001 HEAD CAPACITANCE TESTS: {0.500uF Range} 2.002 DISP Set the Capacitance Decade to 0.470uF. # Note, You must substitute your actual capacitance decade tolerance below. 2.003 ACC 0.470uF 1U 2.004 MEMI Enter UUT reading in microfarads: 2.005 MEME 2.006 MEMC 0.5 uF 1% 0.002U 3.001 HEAD CAPACITANCE TESTS: {0.0500uF Range} 3.002 DISP Set the Capacitance Decade to 0.0470uF. # Note, You must substitute your actual capacitance decade tolerance below. 3.003 ACC 0.0470uF 1U 3.004 MEMI Enter UUT reading in microfarads: 3.005 MEME 3.006 MEMC 0.05 uF 1% 0.0002U 4.001 HEAD CAPACITANCE TESTS: {5.00nF Range} 4.002 DISP Select Relative mode (REL) to zero the display and 4.002 DISP automatically subtract the residual meter and test lead 4.002 DISP capacitance. 4.003 DISP Set the Capacitance Decade to 4.70nF. # Note, You must substitute your actual capacitance decade tolerance below. 4.004 ACC 4.70nF 1U 4.005 MEMI Enter UUT reading in nanofarads: 4.006 MEME 4.007 MEMC 5 nF 1% 0.35U 5.001 DISP Remove all connections. 5.002 HEAD 5.003 END